Lab Quality Control: Levey-Jennings Charts and Westgard Rules
Before a lab reports patient results, it checks that its analyzer is working correctly by testing control samples with known values. This guide explains how control charts work, what each Westgard rule means, and what to do when a rule is broken.
Labs run control samples with known values alongside patient samples and plot the results on a Levey-Jennings chart, with lines at the mean and ±1, 2 and 3 standard deviations (SD). The Westgard rules say when to reject a run: for example, one control beyond 3 SD (1-3s), or two in a row beyond the same 2 SD line (2-2s).
Why Labs Run Quality Control
An analyzer can drift because of a reagent, a lamp, a pipette or temperature. Patients cannot see this. So labs test control samples, which have known values, in each run. If the control results look wrong, the lab stops and fixes the problem before reporting patient results.
The Levey-Jennings Chart
- Run the control many times to find its average (mean) and standard deviation (SD), a measure of normal spread.
- Draw lines at the mean, and at ±1 SD, ±2 SD and ±3 SD.
- Plot each new control result on the chart, in date order.
Most results should sit close to the mean, spread evenly above and below it.
Setting up a lab or clinic?Get a written quote, brochure or demo from the Pulse team.
The Westgard Rules
| Rule | What it means | Action |
|---|---|---|
| 1‑2s | One control beyond ±2 SD | Warning: check the other rules |
| 1‑3s | One control beyond ±3 SD | Reject the run |
| 2‑2s | Two controls in a row beyond the same 2 SD line | Reject the run |
| R‑4s | In one run, one control above +2 SD and another below −2 SD | Reject the run |
| 4‑1s | Four controls in a row beyond the same 1 SD line | Reject the run |
| 10x | Ten controls in a row on the same side of the mean | Reject the run |
The 4-1s and 10x rules usually look back across several runs, because one run has too few control results. The 1-2s line on its own does not reject a run; it tells staff to check the other rules.
What to Do When a Rule Is Broken
A broken rule means the run is rejected, so patient results from it should not be released. We suggest these steps:
- Do not just re-run the control until it passes.
- Look at which rule failed. A single wild result may point to a bubble, clot or pipetting slip. A steady shift to one side often points to a reagent, calibration or lamp problem.
- Check the reagent lot and expiry, the calibration, and the analyzer's maintenance log.
- Fix the cause, recalibrate if needed, and re-run the controls.
- Re-test the patient samples from the failed run, and record what happened.
Let the Analyzer Help
When QC runs in software, the 1-2s warning step is not needed, and the lab can choose rejection rules for each test based on how precise the method is. Analyzers that store QC data and draw the charts save time and reduce mistakes.
For how the tests themselves work, see our how a biochemistry analyzer works and 3-part vs 5-part haematology analyzer guides.
QC on Pulse Analyzers
- Pulse ZT370: four QC modes (L-J, X, X-R and X-B) with manual and automatic calibration.
- Pulse ZT Smart SAA: stores 25,000 test records with reaction curves and QC results.
Lab Quality Control: Key Facts
- Control samples with known values are tested with patient samples in each run.
- A Levey-Jennings chart shows control results against the mean and ±1, 2 and 3 SD.
- The 1-2s rule is a warning; 1-3s, 2-2s, R-4s, 4-1s and 10x reject the run.
- Patient results from a rejected run should not be released.
- In software QC, the lab chooses rejection rules for each test.
Lab Quality Control: FAQs
What are the Westgard rules?
A set of quality control rules used with Levey-Jennings charts: 1-2s as a warning, and 1-3s, 2-2s, R-4s, 4-1s and 10x to decide when to reject a run.
What is a Levey-Jennings chart?
A chart of control results plotted in date order, with lines at the mean and at plus or minus 1, 2 and 3 standard deviations.
What does the 1-3s rule mean?
One control result is more than 3 standard deviations from the mean. The run is rejected.
What does the 10x rule mean?
Ten control results in a row fall on the same side of the mean. This suggests a steady shift, and the run is rejected.
What should a lab do when QC fails?
Hold patient results, find and fix the cause, recalibrate if needed, re-run controls, then re-test the patient samples and record what happened.
Do Pulse analyzers support QC?
Yes. The ZT370 has L-J, X, X-R and X-B QC modes, and the ZT Smart SAA stores QC results. Message +91 90711 01108 for a quote.