Sirolimus-coated PTCA balloon catheter that delivers the drug into the artery wall during angioplasty, without leaving a stent behind. Drug particles of different sizes in a matrix coating help the drug reach each layer of the vessel wall.

Sirolimus 2.75 µg/mm² Semi-Compliant Rapid Exchange
Get a Quote Contact Us Download Brochure

Key Features

Technical Specifications

Product typeSirolimus-coated PTCA dilatation catheter (drug-coated balloon)
Balloon materialNylon
DrugSirolimus
Drug loading2.75 µg/mm²
Balloon typeSemi-compliant
Balloon diameters1.50, 2.00, 2.50, 3.00, 3.50, 4.00 mm
Balloon lengths6, 10, 12, 15, 20, 25, 30, 35, 40 mm
Catheter length142 ± 2 cm
Rapid exchange length25 ± 1 cm
Tip entry profile0.45 ± 0.02 mm
Crossing profile0.60 mm minimum, 0.84 mm maximum
Proximal shaft outer diameter1.97 F (0.653–0.660 mm)
Distal shaft outer diameter2.56 F (0.83–0.87 mm)
Proximal shaft inner diameter0.480 ± 0.01 mm
Nominal pressure6 atm
Rated burst pressure14 atm
Average burst pressureNot less than 18 atm
Minimum guiding catheter5 F (0.056 inch)
Maximum guidewire0.014 inch (0.36 mm)
Inflation timeNot more than 10 seconds
Deflation timeUnder 20 seconds on average
CertificationISO 13485:2016, CE (notified body 1035)

Mamba Sizes and Order Codes

Mamba comes in 6 diameters and 9 lengths. Order codes follow the pattern MD + diameter + length (for example, MD30020 is 3.00 mm × 20 mm).

Length1.50 mm2.00 mm2.50 mm3.00 mm3.50 mm4.00 mm
6 mmMD15006MD20006MD25006MD30006MD35006MD40006
10 mmMD15010MD20010MD25010MD30010MD35010MD40010
12 mmMD15012MD20012MD25012MD30012MD35012MD40012
15 mmMD15015MD20015MD25015MD30015MD35015MD40015
20 mmMD15020MD20020MD25020MD30020MD35020MD40020
25 mmMD15025MD20025MD25025MD30025MD35025MD40025
30 mmMD15030MD20030MD25030MD30030MD35030MD40030
35 mm—MD20035MD25035MD30035MD35035—
40 mm—MD20040MD25040MD30040MD35040—

— Not available.

Mamba Balloon Compliance Chart

Balloon diameter (mm) at each inflation pressure. Green is the nominal pressure (6 atm); red is the rated burst pressure (14 atm). Do not inflate beyond the rated burst pressure.

Pressure (atm)1.50 mm2.00 mm2.50 mm3.00 mm3.50 mm4.00 mm
21.281.772.232.653.13.5
41.41.92.392.863.313.82
6 (nominal)1.522.533.54
81.582.072.573.113.654.13
101.642.142.633.183.764.23
121.692.222.683.243.844.31
14 (rated burst)1.732.292.733.313.94.38
161.792.392.783.373.964.44
181.862.52.823.444.044.48
201.942.642.863.524.144.52

How a Drug-Coated Balloon Like Mamba Is Used

In a drug-coated balloon procedure, the narrowed artery is usually prepared first with a plain angioplasty balloon. The cardiologist then passes the drug-coated balloon to the same spot over a guidewire and inflates it with an inflation device. While the balloon presses against the wall, the drug moves into the tissue. The balloon is then deflated and removed, so no metal is left in the artery.

Mamba carries sirolimus, the same family of drug used on the MedSure and Sirova stents. The choice between a drug-coated balloon and a stent is a clinical decision for the treating cardiologist.

What to Check Before Adding a Drug-Coated Balloon to Your Cath Lab

The Pulse Cath Lab Consumables Range

Pulse supplies the main single-use items for a cath lab, so one supplier answers for supply and documents.

Pulse productWhat it does
AcuSafe Micro-Access Introducer KitsSmall-needle first puncture of the artery or vein, then a sheath
AcuSafe Introducer KitsSheath and dilator that keep the vessel open for catheters and wires
GuideX GuidewiresWires that guide sheaths and catheters into place and support exchanges
Optima ManifoldsControl contrast injection, flushing and pressure monitoring
Glider PTCA balloons (SC, NC, CTO)Open the narrowed artery before a stent, and expand the stent fully after
Optima Inflation Device KitInflates the balloon to a measured pressure to open the artery and expand the stent
Evera, Sirova, EvroSure and MedSure stentsDrug-eluting stents that hold the opened artery open
Mamba drug-coated balloon (this page)Delivers sirolimus to the artery wall without leaving a stent behind
AcuSafe Blunt NeedleControlled access for percutaneous drainage (interventional radiology)

Share the sizes you use and your monthly volume, and request a quote. Ask for the brochure, IFU and certificates with your quote.

Complete the setup

You May Also Need

Buying guides

Related Guides

← Back to Cardiac Care

Got Questions?

Mamba Drug-Coated Balloon FAQs

A drug-coated balloon (DCB) is an angioplasty balloon coated with a drug. When it is inflated in a narrowed artery, the drug passes into the artery wall to help stop it narrowing again. Unlike a stent, nothing is left behind once the balloon is removed.

Mamba is coated with sirolimus at 2.75 µg/mm², in a matrix coating with particles of different sizes so the drug can reach each layer of the vessel wall.

Diameters of 1.50, 2.00, 2.50, 3.00, 3.50 and 4.00 mm, and lengths of 6 to 40 mm. The 1.50 mm and 4.00 mm sizes go up to 30 mm long.

The nominal pressure is 6 atm, the rated burst pressure is 14 atm and the average burst pressure is not less than 18 atm.

Mamba is a rapid exchange catheter that works with a 5F (0.056 inch) or larger guiding catheter and a guidewire up to 0.014 inch.

The Mamba brochure lists ISO 13485:2016 and CE marking (notified body 1035). Ask the Pulse team for the current CDSCO licence and certificates.

Message Pulse on WhatsApp at +91 90711 01108 with your sizes and monthly volume, or download the brochure on this page for full specifications.