November 17, 2026 • Blog • Critical Care • 4 Min Read

Anaesthesia Machine Checklist: What to Check Before Every List

Checking the anaesthesia machine before use is a simple way to keep patients safe. This guide lists the checks the UK Association of Anaesthetists recommends, when to do them, and what to do if the machine fails.

By Pulse Team · Updated November 21, 2026 · 4 min read
Share

Check the anaesthesia machine at the start of every theatre session and before each new patient. Confirm power, gas pipelines and cylinders, test the oxygen failure alarm, check vaporisers, do a breathing system leak test and two-bag test, and check the ventilator, scavenging, suction, monitors and airway kit. A self-inflating bag must always be at hand.

When to Check

These timings come from the 2012 guideline on checking anaesthetic equipment from the UK Association of Anaesthetists (then AAGBI).

The Pre-Use Checklist

CheckWhat to look for
PowerPlugged into the mains and on; backup power ready; no extension leads
Gas supplyEnough oxygen, with a spare cylinder
Pipelines and cylinders"Tug test" each pipeline; gauges at 400 to 500 kPa; no carbon dioxide cylinder fitted
Oxygen failure alarmSounds for at least 7 seconds; know which gases keep flowing
VaporisersLocked in, upright, filled but not too full, dials turn fully, then off
Breathing systemPush and twist each joint; check the CO2 absorber; leak test at 20 to 60 cmH2O
Two-bag testAfter the breathing system, vaporisers and ventilator are each checked
VentilatorSet up right, tubing secure, alarms and relief valve working
ScavengingOn, and connected to the correct port
SuctionWorking, connections tight, suction builds quickly
MonitorsWorking, alarm limits set; check the oxygen, SpO2 and CO2 monitors
Airway equipmentRight sizes ready; new filter and tube mount for each patient; laryngoscopes work; difficult airway kit at hand

Equipping an operation theatre?Get a written quote, brochure or demo from the Pulse team.

Always Have a Backup

A self-inflating bag must be close at hand wherever anaesthesia is given, along with a spare oxygen source. Staff should also know what the machine does if it fails. Some machines fall back to little or no gas flow, or to oxygen with no anaesthetic. So there must be another way to give oxygen, breathe for the patient and keep them asleep.

Training Comes First

The guideline says no one should use the equipment unless they are trained and able to use it. Some machines run a self-test when switched on. We suggest treating the self-test as one part of the check, not the whole check, and following the maker's list for your model.

Monitoring Matters

The CO2 monitor is one of the most useful checks. It helps show the tube is in the airway and the patient is breathing. See our EtCO2 and capnography and patient monitor normal values guides.

Planning a New Theatre?

A safe theatre needs the workstation, piped gases, suction, monitors and airway kit to work together. See our operation theatre equipment list and anaesthesia machine price guides.

The Pulse Anaesthesia Workstation

The Pulse anaesthesia workstation is built for adult and paediatric patients. It has:

Pulse critical care devices are CDSCO cleared and CE marked, made under ISO 13485.

Anaesthesia Machine Checklist: Key Facts

Anaesthesia Machine Checklist: FAQs

How often should an anaesthesia machine be checked?

At the start of every theatre list, before each new patient, after any serious incident, and when a part is changed. Test the oxygen failure alarm each week.

What is the two-bag test?

A final check of the whole breathing system with a test lung or second bag, done after the breathing system, vaporisers and ventilator have each been checked.

What should pipeline gas pressure be?

The UK guideline says pipeline gauges should read 400 to 500 kPa. Do a tug test on each pipeline to make sure it is firmly connected.

How long should the oxygen failure alarm sound?

For at least 7 seconds. Staff should also know which gases keep flowing when oxygen fails.

What backup is needed if the anaesthesia machine fails?

A self-inflating bag and spare oxygen must be close at hand, with another way to keep the patient asleep.

Does Pulse supply anaesthesia workstations?

Yes. The Pulse anaesthesia workstation supports adult and paediatric patients, with two interlocked vaporisers and battery backup. Message +91 90711 01108 for a quote.

Sources

This guide is for general information. Follow your hospital's anaesthesia policy and the workstation maker's checklist. Last reviewed November 2026.