How Does an Autoclave Work? Temperature, Time and Tests
An autoclave kills germs on instruments with hot steam under pressure. This guide explains how it works in simple steps, the standard temperatures and times, gravity vs vacuum autoclaves, and the tests that show a cycle worked.
An autoclave sterilises instruments with steam under pressure. Pressure lets the steam get hotter than boiling water, and that heat kills germs and spores. Standard wrapped cycles, per the US CDC, are 30 minutes at 121°C in a gravity autoclave or 4 minutes at 132°C in a vacuum (prevacuum) autoclave.
How an Autoclave Works, Step by Step
- Load: clean instruments go into the chamber, wrapped or in trays, with space between them.
- Remove air: air must leave the chamber, because trapped air stops steam reaching every surface.
- Heat and hold: steam fills the chamber, the pressure rises, and the load is held at the set temperature for the set time.
- Exhaust: steam is let out and pressure falls.
- Dry and check: packs are dried, then indicators are checked before use.
The CDC lists four things that must be right: steam, pressure, temperature and time.
Autoclave Temperature and Time
| Autoclave type | Temperature | Minimum time (wrapped items) |
|---|---|---|
| Gravity displacement | 121°C (250°F) | 30 minutes |
| Prevacuum (vacuum) | 132°C (270°F) | 4 minutes |
These are the minimum exposure times in the CDC guideline. Times change with the load, wrapping and machine, so always follow the maker's instructions. Waste needs longer: the CDC notes that 10 lb (about 4.5 kg) of lab waste needs at least 45 minutes at 121°C.
Gravity vs Vacuum Autoclaves
| Gravity autoclave | Prevacuum (vacuum) autoclave | |
|---|---|---|
| How air is removed | Steam pushes air out through a drain | A vacuum pump pulls air out first |
| Best for | Unwrapped metal items, liquids, lab media, waste | Wrapped packs, porous loads, hollow items |
| Speed | Slower | Faster; steam reaches inside packs almost at once |
Vacuum autoclaves are often called Class B. See our autoclave price guide for what each type costs.
Choosing an autoclave?Get a written quote, brochure or demo from the Pulse team.
How to Check That a Cycle Worked
| Check | How often | What it shows |
|---|---|---|
| Gauges or printout (time, temperature, pressure) | Every load | The machine reached its settings |
| Chemical indicator inside each pack | Every pack | Steam reached the inside of the pack |
| Bowie-Dick test (vacuum autoclaves) | Daily, before the first load | Air is being removed properly |
| Biological indicator (spore test) | At least weekly, and every load with implants | Spores were killed |
If a chemical indicator does not change colour, do not use the instruments. If a spore test is positive, take the autoclave out of service, hold any implants from that load and repeat the test. These steps follow CDC guidance.
Common Mistakes
- Overloading the chamber, so steam cannot reach every item.
- Skipping cleaning first. Dirt shields germs from steam.
- Opening the door early, so packs come out wet.
- Using a gravity cycle for wrapped or hollow items it is not meant for.
- Not keeping a log of cycles and test results.
Which Autoclave for Which Facility
- Clinics and dental chairs: portable or dental autoclaves, 12 to 23 litres.
- Nursing homes and small hospitals: vertical autoclaves, 40 to 180 litres.
- Hospital CSSD and OT complexes: horizontal autoclaves for large drums and trays.
See also our OT equipment list and modular OT setup cost guides.
Pulse Autoclaves and Sterilizers
Pulse supplies autoclaves from 12 to 180 litres: portable, vertical, horizontal and dental models, in stainless steel or aluminium, with electric and non-electric options. Sizes can be customised for your project.
How an Autoclave Works: Key Facts
- An autoclave sterilises with steam under pressure; steam, pressure, temperature and time must all be right.
- CDC minimum for wrapped items: 30 minutes at 121°C (gravity) or 4 minutes at 132°C (prevacuum).
- Vacuum autoclaves remove air with a pump, so steam reaches inside wrapped and porous loads.
- Vacuum autoclaves need a daily Bowie-Dick test before the first load.
- Spore tests are done at least weekly and with every load containing implants.
How an Autoclave Works: FAQs
How does an autoclave work?
It removes air from the chamber, fills it with steam under pressure, holds the load at a set temperature for a set time, then lets the steam out and dries the load.
What temperature and time does an autoclave use?
Per CDC guidance, wrapped items need at least 30 minutes at 121°C in a gravity autoclave, or 4 minutes at 132°C in a prevacuum autoclave.
What is the difference between a gravity and a vacuum autoclave?
A gravity autoclave lets steam push air out. A vacuum (prevacuum or Class B) autoclave pumps air out first, so steam reaches inside wrapped packs and hollow items.
What is a Bowie-Dick test?
A daily test for vacuum autoclaves, run before the first load, that checks whether air is being removed from the chamber properly.
How often should an autoclave be spore tested?
At least weekly, and with every load that contains implants, using a biological indicator.
Does Pulse supply autoclaves?
Yes. Pulse supplies portable, vertical, horizontal and dental autoclaves from 12 to 180 litres. Message +91 90711 01108 for a quote.
Sources
- Steam sterilization (US CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008; last reviewed 28 November 2023): cycle times, four parameters, gravity vs prevacuum, Bowie-Dick test, waste
- Best practices for sterilization monitoring (US CDC, last reviewed 15 May 2024): mechanical, chemical and spore test monitoring; positive spore test steps
- Glossary (US CDC): autoclave, biological indicator, temperature equivalents
Related Reading